Product Description
Features:
Portable handle, all-in-one probe holder.
12.1 inch foldable color TFT LCD Screen, display waveform and data simultaneously.
FHR, TOCO, FM and waveform of fetal activities can be monitored.
Count fetal movement in auto/manual mode.
The color of display area can be adjustable.
It can store at least 24hrs CTG data in power-down memory. Data won’t be erased after switch off. Recall printing is available.
Alarm review.
Auto/manual patient ID generation, easy user-management & data searching.
25mm/s high speed review of printing.
Alarm light on the top. Smart alarm setting: light & sound alarm, alarm range and tone can be adjusted. It also owns Sensor off alarm, monitor abnormal alarm and remark function.
Two type of printing function: American standard and National standard.
FHR2 offset printing function.
Three monitoring and display mode: Fetal, Fetal & Maternal, Maternal.
Real-time printing and timing printing.
Event remark.
Built-in high capacity rechargeable battery.
Support paper-out cache, selectable printing and auto shutdown.
Independent network communicate function, support remote, wireless, wired networking function.
It can be upgraded to maternal multi-parameter monitoring, it has various of bracket solution
Optional configuration Twin FHR, Fishers auto fetal-analysis, fetal awake-up alarm, wireless network, wall mount, moving cart, 12 volts Power supply
Specifications
FHR
Ultrasound Frequency: 1MHz±1KHz
Ultrasound input intensity: < 10mW/c㎡
Range: 30~240bpm
Accuracy: ±1bpm
Alarm limit & tolerance: Max: (30 ~ 240)bpm, tolerance: ±5%.
Min: (30 ~ 240)bpm, tolerance: ±5%
TOCO
Measurement range: 0~100%
Resolution: 1%
Pressure tolerance: +6.7%
ECG:
Lead Selection: I、II、III、aVR、aVL、aVF、V
Heart Rate Range: 15-300bpm
Accuracy: ±1bpm or ±1%, consider the larger one.
Sweep speed: 6.25 mm/s 、12.5 mm/s 、25 mm/s、50mm/s
Gain Selection: ×1/4 、×1/2、×1、×2、×4、Auto
Calibration: ±1 mv
NIBP:
Philosophy: Oscillometric method
Measurement range: Systolic 30-270mmHg Diastolic 10-200mmHg
Mean 20-220mmHg Measurement mode: Manual/Automatic/continuous measurement
Auto-measurement cycle: 0-480mins adjustable
Other function: NIBP Trend tabulation, supplementary venipuncture function.
SPO2
Display type: Value, waveform, plethysmograph
Measurement range: 0-100%
Accuracy: 70%-100%(±2%);50%-69%(±3%);0%-49% (undefined)
Pulse Rate
Measurement Range: 30-250bpm
Accuracy: ±1bpm or ±2%, consider the larger one.
Respiration
Philosophy: Thoracic impedance & nasal tube
Measurement range: 0-120bpm
Accuracy: ±1bpm or ±5%, consider the larger one.
Lead Selection: RA-LA,RA-LL,LA-LL
Temperature
Number of channel: 2channerls, display T1、T2、△T
Measurement range: 25℃~45℃
Accuracy: ±0.2℃
Sensor: Standard config with skin temp sensor.